Over 300 New Jersey Auctions End Tomorrow 05/08 - Bid Now
Over 100 Total Lots Up For Auction at One Location - WA 05/09

spacer
Listing 4657631 Listing 4657631 Listing 4657631 Listing 4657631 Listing 4657631 Listing 4657631 Listing 4657631 Listing 4657631
Listing 4657631 Listing 4657631 Listing 4657631 Listing 4657631 Listing 4657631 Listing 4657631 Listing 4657631 Listing 4657631

For Sale SYSMEX RD-100i One Step Nucleic Acid Analyser Detector OSNA CK19 mRNA RT-Lamp Amylase/Lipase Analyzer

  • Asking Price :£2,850.00 GBP [convert]
  • Condition : Used - Excellent
  • Qty. Available : 1
  • In Stock : Yes
  • Date : May 07, 2024
  • Listing # : 4657631
Advertisement

Item Description

Seller assumes all responsibility for this listing..

Listing: #4657631

  • Condition: Used - Excellent
  • Brand: SYSMEX
  • Type: Amylase/Lipase Analyzer
  • Model: RD-100i One Step Nucleic Acid Analyser Detector OSNA CK19 mRNA RT-Lamp
  • MPN: RD-100i
Sysmex RD-100i One Step Nucleic Acid Analyser Detector OSNA CK19 mRNA RT-Lamp

In Good Condition And Full Working Order

Included:-
1 x Sysmex RD-100i One Step Nucleic Acid Analyser Detector OSNA CK19 mRNA RT-Lamp
1 x HP Monitor
1 x HP RP5800 i3 3.3GHz PC - Preloaded with RD Software
1 x Keyboard & Mouse
1 x HP LaserJet 1320 Printer
1 x Sysmex RD-100i Stand
1 x Sysmex Test Protocol Manual
1 x Sysmex RD-100i Software User Guide
1 x Set of Setup Software DVD


The test procedure
The lymph node tissue is homogenised in one simple step prior to analysis. Further filtration of the RNA is not required. The samples are then processed and measured under standardised conditions in the RD-100i, an automated real time analyser. Up to four lymph nodes (up to 8 samples for colon cancer) can be analysed in parallel. The results are available in approximately 30 minutes.
The device features a 16-minute amplification reaction in real time. Depending on the number of measured copies of the mRNA marker, the results are shown in one of three categories:
(++) = Indication for a macrometastatic tumour burden
(+) = Indication for a micrometastatic tumour burden
(–) = No metastatic disease

Advantages of the OSNA® method:
Simple sample processing
Isothermal reaction at 65°C
Very fast amplification reaction (16 minutes)
Indication of the size of the metastatic lesions

Technical specifications
Instrument - Gene Amplification Detector System RD-100i
Method - OSNA® (One Step Nucleic Acid Amplification)
Parameter - CK19 mRNA
Technology - RT-LAMP (Reverse transcriptase loop-mediated isothermal amplification)
Detection - Change of transmitted light caused by the precipitation of magnesium
pyrophosphate depending on the grade of the reaction
Displayed parameters - CK19 Q (CK19 Qualitative Result)
CK19 (CK19 Risetime)
CK19 C (CK19 mRNA-concentration)
Throughput - 4 samples per batch
Tissue range per sample - 50–600 mg
Sample volume - 2 μL
Reagents - homogenising reagent LYNORHAG
amplification reagent LYNOAMP BC
all reagents are ready for use
Data storage - 2000 samples
Quality control - positive and negative control
180 data points per file
Interfaces - Host computer connection (RS232, LAN)
printer connection (USB)
Weight - approx. 66 kg
Dimensions (W x H x D) - 596 x 548 x 622 mm

Return Policy

Items are sold as-is with no returns or refunds available unless explicitly stated.
Seller Information
SaveOnKit
Shop with confidence
Citi Capital Groupspacer
Send the Seller a message: