UP-7000
Multi-parameter Patient Monitor
NIBP
Technique Oscillometric
Typical measuring time <30 seconds (typical adult cuff)
Initial cuff inflation pressure Adult <175mmHg
Pediatric <135mmHg
Neonate <65mmHg
Overpressure protection limit Adult 300mmHg
Pediatric 240mmHg
Neonate 150mmHg
Measuring range
Systolic pressure Adult 40mmHg~275mmHg
Pediatric 40mmHg~200mmHg
Neonate 40mmHg~135mmHg
Diastolic pressure Adult 10mmHg~210mmHg
Pediatric 10mmHg~150mmHg
Neonate 10mmHg~95mmHg
Mean arterial pressure Adult 20mmHg~230mmHg
Pediatric 20mmHg~165mmHg
Neonate 20mmHg~110mmHg
Measurement accuracy Maximum mean difference: ±5 mmHg
Maximum standard deviation: 8 mmHg
Measurement mode Manual, Auto, STAT
Automatic measuring intervals 1~480min
TEMP
Measuring range 21.0°C~50.0°C
Measuring accuracy ±0.2°C for range from 25.0°C~45.0°C
SpO2
Transducer Dual-wave length LED
SpO2 measuring range 0%~100%
SpO2 measuring accuracy 2% for range from 70% to 100%
Low perfusion performance As low as 0.4%
PR measuring range 0bpm~250bpm
PR measuring accuracy ±2bpm or ±2%, whichever is greater
ECG
Input dynamic range ±0.5mVp~±5mVp
HR measuring range 15bpm~350bpm
HR measuring accuracy ±1% or ±2bpm, whichever is greater
HR alarm delay time ≤10s
Sensitivity selection ×1/4, ×1/2, ×1, ×2, ×4 and Auto
Sweeping speed 6.25mm/s,12.5mm/s, 25mm/s, 50mm/s
ECG noise level ≤30μVp-p
ECG input loop current ≤0.1μA
Differential input impedance ≥10Mohm
Common-mode rejection ratio ≥105dB (Monitoring mode)
(CMRR) ≥89dB (Diagnostic mode)
Time constant ≥0.3s (Monitoring mode)
≥3.2s (Diagnostic mode)
RESP
RR measuring range 0rpm~120rpm
RR measuring accuracy ±5% or ±2rpm, whichever is greater
Others
Power supply 100~240Vac, 50/60Hz
Display 12.1 inch TFT display
Alarming mode Audible-visual alarm
Networking port Ethernet port
Standard Configuration
ECG, RESP, SpO2, NIBP, TEMP, PR
Option
2-IBP, EtCO , Nellcor SpO , SunTech NIBP, Cardiac Output, Built-in printer,
Cerebral State Monitoring, Central monitoring system, Touch screen
Return Policy
Items are sold as-is with no returns or refunds available unless explicitly stated.
Tags: