Awareness Technologies ChemWell 2902 Chemistry Analyzer
ChemWell 2902 Automated PC controlled biochemistry analyzer
- Fully automated dilutions capability, pre-dilutions, dispense reagents by single well or 8-way multiple well reagent dispense
- Reactions take place in standard 96 well microwell plates or microwell strips
- Maximum number of specimens: 96 (including calibrators and controls)
- Maximum number of reagents: Typically 27 or 44 (reagents can also be programmed to go to the sample rack)
- Automated plotting and editing reaction graphs, calibration curves, including kinetic reactions
- Maximum throughput of 200 tests per hour (test dependent) QC Tracking options, using Levey-Jennings
- 8 filter wheel (340-700 nm)
- Applications in clinical and veterinary testing; environmental testing; analysis of food and water; and life science research
- Optional Reagent Cooling Accessory (RCA)
Specifications
Typical throughput up to 200 endpoint reactions per hour, up to 170 kinetic reactions per hour
Dimensions 34″ (86cm) left to right, 16″ (40 cm) height, 20″ (51 cm) front-to-back
Capabilities Dilution, pre-dilution, dispensing single or multiple reagents
Probes 316 stainless steel for maximum reagent compatibility, level sensing
Maximum number of specimens 96
Maximum number of reagents Typically 27 to 44 or more (you can program reagents to go in the
sample rack; Assorted replaceable racks and custom designed racks
are available for various bottle sizes)
Standard reagent rack 27
Reaction Vessel Standard micro wells, strips or plates
Instrument bottles 1L priming bottle
Thermal control Well, probe, and tubing; ambient or 37C (other options also available)
Cooling control RCA, Reagent Cooling Accessory optional) cools 12 to 15C below ambient
through peltier thermoelectric modules connected to an external controller
Optical design Reads absorbance in 4 simultaneous channels, NIST traceable calibration
user selected monochromatic or biochromatic results 8 position filter wheels:
340, 405, 450, 505, 545, 600, 630, 700 or custom
Interference filters Long life, hard coat, ion assisted deposition, 10nm typical half bandpass
Linear Range 0.2 to 3.0A
Photometer Accuracy +/- 1% or better
Certifications NRTL Listed, CE Marked
Return Policy
Items are sold as-is with no returns or refunds available unless explicitly stated.