Item Description

Seller assumes all responsibility for this listing..

Listing: #2899293

  • Condition: Used - Good
  • Brand: ALCON
  • Type: Phacoemulsifier
  • Model: INFINITI
  • Power Supply Voltage: 90-250
  • Year of Manufacture: 2009
  • Warranty: 1 year
  • Software Version: 2.06
This is a 2009 system that has been fully refurbished and has the latest 2.06 software. The system comes with one OZil handpiece and remote plus operators manual. If you would like a short video of the system and additional details, please do not hesitate to ask. We are a one stop shop for eye surgical equipment and are experts with phaco and vitret systems. The system is guaranteed for one year and if you want a discount, we can reduce the warranty period for you.

We will also teach you how to do your own PM on the system so that you can same money year over year on the expense from Alcon Tech Service. Alcon Tech Service will charge a fee to inspect the equipment prior to doing service. This is a recent change and is designed to create additional revenue for Alcon. Expect Alcon to request a 10,000.00 Purchase Order prior to servicing any Alcon Product that has missed a annual preventative maintenance or has been moved to a new facility.

If the INFINITI system has AQUALASE enabled, chances are it has not seen service in awhile. Ask the broker to turn on the system and select ALCON SETTINGS as the doctor and then press the handpiece type in the upper center of the screen. If Aqualase shows up, then it has been awhile since the system has been service. You will need to have a service rep go into the software set up and disable the AQUALASE feature. You will need to replace the CPU 3 volt battery and the 12 Volt back up battery. Also, the solenoid spacers will need to be replaced so that the system will not go into ERROR mode

Alcon Infiniti Error Code 131 Warning Warning xxx: Fluidics not available. Restart system. If problem persists after restarting system, contact Alcon Technical Services. "Irrigation valve failed closed. Irrigation valve failed open." "• Fluidics Controller PCB Irrigation Valve / Replace Fluidics Module"

Alcon Infiniti Error Code 132 Warning Warning xxx: Fluidics not available. Restart system. If problem persists after restarting system, contact Alcon Technical Services. Vent valve failed closed. Vent valve failed open. "• Fluidics Controller PCB Vent Valve / Replace Fluidics Module"

Alcon Infiniti Error Code 135 Warning Warning xxx: Fluidics not available. Restart system. If problem persists after restarting system, contact Alcon Technical Services Vent valve failed closed. Vent valve failed open. "• Fluidics Controller PCB Vent Valve / Replace Fluidics Module"

Alcon Infiniti Error Code 136 Warning Warning xxx: Fluidics not available. Restart system. If problem persists after restarting system, contact Alcon Technical Services "Pump direction failure. Pump speed failure." "• Fluidics Controller PCB• Pump Motor / Replace"

Alcon Infiniti Error Code 139 Warning Warning xxx: Fluidics not available. Restart system. If problem persists after restarting system, contact Alcon Technical Services "Pump direction failure. Pump speed failure." "• Fluidics Controller PCB• Pump Motor / Replace"

Alcon Infiniti Error Code 140 Warning Warning xxx: Fluidics not available. Restart system. If problem persists after restarting system, contact Alcon Technical Services "APS s/w error. APS offset failure. APS shunt calibration failure. APS actuator failure." "• Fluidics Controller PCB
• Aspiration Pressure Sensor (APS) / Replace Fluidics Module"

Alcon Infiniti Error Code 142 Warning Warning xxx: Fluidics not available. Restart system. If problem persists after restarting system, contact Alcon Technical Services "APS s/w error. APS offset failure. APS shunt calibration failure. APS actuator failure. IPS s/w error.
FMS s/w error." "• Fluidics Controller PCB• Aspiration Pressure Sensor (APS) / Replace Fluidics Module"

Alcon Infiniti Error Code 143 Warning Warning xxx: Fluidics not available. Restart system. If problem persists after restarting system, contact Alcon Technical Services"APS s/w error. APS offset failure. APS shunt calibration failure. APS actuator failure. IPS s/w error.
FMS s/w error." "• Fluidics Controller PCB• Aspiration Pressure Sensor (APS) / Replace Fluidics Module"

Alcon Infiniti Error Code 144 Warning Warning xxx: Fluidics not available. Restart system. If problem persists after restarting system, contact Alcon Technical Services"APS s/w error. APS offset failure. APS shunt calibration failure. APS actuator failure. IPS s/w error. FMS s/w error." "• Fluidics Controller PCB• Aspiration Pressure Sensor (APS) / Replace Fluidics Module"

Return Policy